-40%

Main Board Dv-ti-2000-main for Ningbo David Ti-2000

$ 158.4

Availability: 100 in stock
  • Country/Region of Manufacture: China
  • Model: Ti 2000
  • Brand: Ningbo David
  • Condition: New
  • All returns accepted: ReturnsNotAccepted

    Description

    Main Board DV-TI-2000-MAIN V3.0 p/n T02.008.00 for Ningbo David TI-2000
    All audible alarms can be canceled for 5 minutes through press reset key except for the power trouble alarm and system trouble alarm. But the alarm indication light is still on, moreover, you can not cancel the alarm and indication caused from other reasons when you cancel the alarm. The incubator will continue alarming until the alarm condition is corrected. If in corrected within 5 minutes, the incubator will alarm again
    RAM failure alarm inside of MCU:All alarm indicators is displayed along with a continuous audible alarm to indicate that the incubator has experienced an internal failure. And Skin Temp.¿ display indicates E11.
    System Failure Alarm:All alarm indicators is displayed along with a continuous audible alarm to indicate that the incubator has experienced an internal failure. as well as Skin Temp.¿ display indicates E12.
    ROM failure alarm inside of MCU:All alarm indicators is displayed along with a continuous audible alarm to indicate that the incubator has experienced an internal failure. And Skin Temp.¿ display indicates E25